0%

Philips Respironics Trilogy 200 Ventilator

Features

  • Trilogy provides respiratory support to meet your prescribed needs, up to 24 hours a day
  • Trilogy is designed with easy-to-read and navigate screens, multiple Rx settings, and on-screen waveforms to help optimize treatment.
  • With no inspiratory effort required to trigger a breath, our unique ‘kiss’ trigger detects when a patient engages and disengages from the mouthpiece to deliver on-demand ventilation with the security of patient alarms
  • Digital Auto-Trak’s capability contributes to the patient’s synchrony and comfort, without requiring manual adjustments. The Sensitive Auto-Trak setting provides an enhanced triggering response for patients with minimal respiratory effort
  • From power options and circuit adapters to reporting software and alarms, Trilogy’s array of accessories helps give patients and families more freedom and mobility while maintaining effective treatment.

Description

The Philips Respironics Trilogy 200 portable ventilator is lightweight and easy to use. The Trilogy 200 was designed for the home, hospital, and alternative care sites to provide invasive and noninvasive ventilator support for a wide range of pediatric and adult patients. Compared to the Trilogy 100, the Trilogy 200 has all the features of the 100, plus extra triggering sensitivity utilizing a proximal flow sensor. The extra triggering sensitivity decreases the work of breathing that provides increased comfort

Dimensions

  • Height: 9.5” (23.52 cm)
  • Width: 11.2” (28.45 cm)
  • Depth: 6.6” (16.68 cm)
  • Weight: 11 lbs (5 kg)

Controls

  • Circuit types: Active Flow; Active with PAP; Passive
  • Ventilation types: Pressure and Volume
  • Volume modes: AC; SIMV (w/PS); CV; AC-MPV
  • Pressure modes: CPAP; S; S/T; T; PC-SIMV (w/PS); PC; PC-MPV; AVAPS-EA
  • IPAP: 4 to 50 cm H2O
  • PEEP/EPAP for active circuits: 0 to 25 cm H2O
  • PEEP/EPAP for passive circuits: 4 to 25 cm H2O
  • CPAP for passive circuits: 4 to 20 cm H2O
  • Pressure support differential: 0 to 40 cm H2O

General

  • Tidal volume: 0 to 2000 ml
  • Breath rate: 0 to 60 BPM for AC mode; 1 to 60 BPM for all other modes
  • Inspiratory time: 0.3 to 5.0 sec
  • Rise time: 1 to 6
  • Ramp start pressure:
    • 0 to 25 cm H2O for active circuits
    • 4 to 25 cm H2O for passive circuits
    • 4 to 19 cm H2O in CPAP
  • Ramp Length: Off; 5 to 45 min
  • Flex: Off; 1 to 3
  • Flow Trigger sensitivity: 1 to 9 l/min
  • Flow Cycle: 10 to 90 %
  • Apnea Rate: 4 to 60 BPM
  • Apnea Rate: 4 to 60 BPM

Measured Parameters

  • Respiratory Rate: 0 to 80 BPM
  • Minute Ventilation: 0 to 99 l/min
  • Vte/Vti: 0 to 2000 ml
  • Leak Rate: 0 to 2000 l/min
  • Peak Inspiratory Flow: 0 to 2000 l/min
  • Peak Inspiratory Pressure: 0 to 99 cm H2O
  • Mean Airway Pressure: 0 to 99 cm H2O
  • % Patient Triggered Breaths: 0 to 100 %
  • I: E ratio: 9.9:1 to 1:9.9